Machine technical sheet


Brand: STIHL Andreas AS AG
Model: TS 400
Type: Stone saws
Built in: n.d.
Power supply: Internal combustion gasoline engine
Anti-vibration devices: Absent
Source of the data: Acquisito sul campo da misuratore

Immagine rappresentante la tipologia
Declared values as set forth in the standard EN ISO 19432
( x 1,0 Cutting stone, asphalt, steel and other solid materials )
CONDITIONS MATERIAL ACCESSORY VALUE K(1) Note
No declared data
(1) Expanded uncertainty

Measurements in the field (Click to view the measurements in the field)

Reference person: Alessandro Lunghi ISPESL INAIL (a.lunghi@inail.it) on null
Features of the accessory: NOT ASSESSED
domDebug Array ( [id] => 1319 [data] => [referente] => 58 [dichiarazione_di_pubblicabilita] => Yes [luogo] => [comparto] => Costruzioni stradali e ferroviarie [numero_di_serie] => [accessorio_usato] => 1045 [t_lavoro_effettuato] => 24 [materiale_lavorato] => 195 [leq] => [note] => [foto1] => No [foto2] => No [foto3] => No [pressione] => [frequenza] => [isValid] => Yes [referente2] => 51 [caratteristicheAccessorio] => [durata] => [en_luogo] => [importato] => [caratteristicheLavoroEffettuato] => [caratteristicheMateriale] => Ferro [en_note] => [id_anagrafica_generale] => 959 [owner_user] => 51 [stato] => valida [id_modello_strumento_di_misura] => 20 [id_sensore_strumento_di_misura] => 12 [id_adattatore_del_sensore] => 4 [accessorio] => NON INDICATO [en_accessorio] => NOT ASSESSED [materiale] => Metallo: leghe metalliche dure o soffici [en_materiale] => Metal: Hard or soft metal alloys [denominazione] => ISPESL INAIL [nome_cognome] => Alessandro Lunghi [tel] => 0694181254 [email] => a.lunghi@inail.it [username] => alessandro [categoria] => 2 [lavoro] => Taglio [en_lavoro] => Cutting [en_comparto] => Construction of roads and railways )

Values related to the vibration worst case handle ​ ( Right )
A hx (Mean)
2.7 m/s2
A hy (Mean)
3 m/s2
A hz (Mean)
3.6 m/s2
A hv sum
5.4 m/s2
Standard deviation
0.1 m/s2
Standard deviation
0.1 m/s2
Standard deviation
0.1 m/s2
Standard deviation x 1,645:
0.16 m/s2
Mean
+ Standard deviation:
2.8 m/s2
Mean
+ Standard deviation:
3.1 m/s2
Mean
+ Standard deviation:
3.7 m/s2
A hv sum
+ (Dev. std. x 1,645):
5.6 m/s2
logo regione toscana
Regione Toscana
Diritti Valori Innovazione Sostenibilità
collegamento sito usl 7 siena Servizio Sanitario
della
Toscana
LOADING TIME: 0.023 s.
XS SM MD LG XL

info@portaleagentifisici.it