Machine technical sheet


Brand: Kavo Italia srl (GE)
Model: Intramatic Lux 2-20LN
Type: Dental drill (dentist)
Built in: n.d.
Power supply: Pneumatic
Anti-vibration devices: Present
Source of the data: Acquisito sul campo da misuratore

Immagine rappresentante la tipologia
Declared values as set forth in the standard EN ISO 5349-1: 2004 - EN ISO 5349-2: 2004
CONDITIONS MATERIAL ACCESSORY VALUE K(1) Note
No declared data
(1) Expanded uncertainty

Measurements in the field (Click to view the measurements in the field)

Reference person: Alessandro Lunghi ISPESL INAIL (a.lunghi@inail.it)
Location: Roma on 22/02/2010
Features of the accessory: Milling cutter
domDebug Array ( [id] => 1286 [data] => 2010-02-22 [referente] => 58 [dichiarazione_di_pubblicabilita] => Yes [luogo] => Roma [comparto] => Sanita' [numero_di_serie] => [accessorio_usato] => 1051 [t_lavoro_effettuato] => 33 [materiale_lavorato] => 206 [leq] => [note] => [foto1] => Yes [foto2] => [foto3] => [pressione] => [frequenza] => [isValid] => Yes [referente2] => 58 [caratteristicheAccessorio] => Punta per uso odontoiatrico [durata] => 60 [en_luogo] => [importato] => [caratteristicheLavoroEffettuato] => Rifinitura protesi dentarie [caratteristicheMateriale] => Gesso [en_note] => [id_anagrafica_generale] => 21633 [owner_user] => 51 [stato] => valida [id_modello_strumento_di_misura] => 2 [id_sensore_strumento_di_misura] => 4 [id_adattatore_del_sensore] => 10 [accessorio] => Fresa [en_accessorio] => Milling cutter [materiale] => Gesso [en_materiale] => Gypsum [denominazione] => ISPESL INAIL [nome_cognome] => Alessandro Lunghi [tel] => 0694181254 [email] => a.lunghi@inail.it [username] => alessandro [categoria] => 2 [lavoro] => Fresatura [en_lavoro] => Milling [en_comparto] => Health service )
Conditions of measurement

Values related to the vibration worst case handle ​ ( Right )
A hx (Mean)
0.3 m/s2
A hy (Mean)
0.4 m/s2
A hz (Mean)
0.2 m/s2
A hv sum
0.5 m/s2
Standard deviation
0.1 m/s2
Standard deviation
0.1 m/s2
Standard deviation
0.1 m/s2
Standard deviation x 1,645:
0.16 m/s2
Mean
+ Standard deviation:
0.4 m/s2
Mean
+ Standard deviation:
0.5 m/s2
Mean
+ Standard deviation:
0.3 m/s2
A hv sum
+ (Dev. std. x 1,645):
0.7 m/s2
logo regione toscana
Regione Toscana
Diritti Valori Innovazione Sostenibilità
collegamento sito usl 7 siena Servizio Sanitario
della
Toscana
LOADING TIME: 0.019 s.
XS SM MD LG XL

info@portaleagentifisici.it