Verify that the following actions have been implemented:
creation of a LASER controlled area delimited by protective guards to prevent involuntary access to the LASER radiation, connected to the interlock system;
display of hazard warning signs, indicating class and type of emitted radiation, at each entrance to the LASER controlled area;
presence of the personal eye-protection equipment recommended by the manufacturer and by the LASER Safety Officer;
presence of labels, indicating class and type of emitted radiation, on the LASER and on the manual;
written procedures covering the correct use;
written procedures covering the alignment (if required as part of the maintenance by the operators);
training and information for the operators working on the device;